AP Chemistry · Topic 3.13 · Reading guide, about 5 min

Using absorbance to find concentration

Use the Beer-Lambert law A = epsilon b c, calibration curves and good spectrophotometer technique to find the concentration of a colored solution.

In one minute

A colored solution takes in some of the light that passes through it. The more concentrated the solution, the more light it takes in. A spectrophotometer measures this absorbance, and because absorbance is proportional to concentration, a chemist can turn a reading into a concentration.

This topic ties together light absorption, linear graphs and careful lab technique, and it often appears in lab-based free-response questions.

Key terms

Absorbance (A)
A unitless measure of how much light a sample absorbs at one wavelength. A higher absorbance means less light gets through.
Beer-Lambert law
The relationship A=ϵbcA = \epsilon b c: absorbance is proportional to path length and to concentration.
Molar absorptivity
The constant ϵ\epsilon that says how strongly a substance absorbs at a given wavelength; its units are L mol−1cm−1\text{L mol}^{-1}\text{cm}^{-1}.
Path length
The distance bb the light travels through the sample, set by the width of the cuvette, usually 1.00 cm.
Cuvette
A small, clear container, often a square tube, that holds the sample in the spectrophotometer.
Spectrophotometer
An instrument that sends light of a chosen wavelength through a sample and measures how much is absorbed.
Wavelength of maximum absorbance
The wavelength where a substance absorbs most strongly. Measurements there give the largest, most sensitive signal.
Calibration curve
A graph of absorbance against concentration for standard solutions, used to read the concentration of an unknown.
Blank
A cuvette holding only the solvent, used to set the instrument to zero absorbance before measuring samples.

The Beer-Lambert law

The law says A=ϵbcA = \epsilon b c, where AA is absorbance, ϵ\epsilon is molar absorptivity, bb is path length in centimeters and cc is molar concentration. Doubling the concentration doubles the absorbance, and doubling the path length does too. Plotting absorbance against concentration for a fixed cuvette gives a straight line through the origin with slope ϵb\epsilon b.

The law holds well at modest concentrations. At very high concentrations the line bends and the law stops being reliable, so a very dark sample is diluted before measuring. Exam data usually stay in the straight-line range, and you may be asked why a measurement outside that range cannot be trusted.

Absorbance has no unit, so ϵ\epsilon must carry units that cancel with bb and cc: L mol−1cm−1×cm×mol L−1\text{L mol}^{-1}\text{cm}^{-1} \times \text{cm} \times \text{mol L}^{-1} leaves nothing.

Choosing the wavelength and building the curve

First, scan a sample across wavelengths and find where the absorbance peaks. Work at that wavelength, because the signal changes most for a small change in concentration, and small errors in setting the wavelength matter least near the top of a peak. Then prepare several standard solutions of known concentration, zero the instrument with a blank, and record each absorbance.

Plot absorbance on the vertical axis and concentration on the horizontal axis, and draw the best-fit line. To find an unknown, measure its absorbance in the same cuvette at the same wavelength, locate that absorbance on the line, and read across to the concentration axis. The unknown should fall within the range of the standards. Extending the line beyond the data is a guess.

Good technique and what goes wrong without it
StepWhy it mattersTypical error
Use the wavelength of maximum absorbanceHighest sensitivityShallow slope and less precise results
Zero with a blankRemoves absorption by solvent and cuvetteAll readings shifted
Wipe the cuvette and avoid fingerprintsSmudges scatter or absorb lightAbsorbance too high
Rinse the cuvette with the sampleLeftover liquid dilutes or contaminatesLeftover water dilutes the sample, so absorbance and the calculated concentration come out too low
Same cuvette and wavelength for all readingsKeeps b and epsilon fixedPoints off the line

Worked example: an unknown from a standard

Problem: A standard solution of 4.00×10−54.00 \times 10^{-5} M dye gives A=0.480A = 0.480 in a 1.00 cm cuvette. A sample of unknown dye solution, measured under identical conditions, gives A=0.312A = 0.312. The unknown was made by diluting 5.00 mL of the original solution to 50.00 mL. Find the concentration of the original solution and the moles of dye in the 5.00 mL.

Step 1, find ϵ\epsilon. ϵ=A/(bc)=0.480/[(1.00 cm)(4.00×10−5 M)]=1.20×104 L mol−1cm−1\epsilon = A / (bc) = 0.480 / [(1.00\ \text{cm})(4.00 \times 10^{-5}\ \text{M})] = 1.20 \times 10^{4}\ \text{L mol}^{-1}\text{cm}^{-1}.

Step 2, find the diluted concentration. c=A/(ϵb)=0.312/[(1.20×104)(1.00)]=2.60×10−5 Mc = A / (\epsilon b) = 0.312 / [(1.20 \times 10^{4})(1.00)] = 2.60 \times 10^{-5}\ \text{M}.

Step 3, undo the dilution. The dilution factor is 50.00/5.00=10.050.00 / 5.00 = 10.0, so the original concentration is 2.60×10−5×10.0=2.60×10−4 M2.60 \times 10^{-5} \times 10.0 = 2.60 \times 10^{-4}\ \text{M}.

Step 4, moles in 5.00 mL. (2.60×10−4 mol/L)(0.00500 L)=1.30×10−6 mol(2.60 \times 10^{-4}\ \text{mol/L})(0.00500\ \text{L}) = 1.30 \times 10^{-6}\ \text{mol}. Check: the unknown absorbance is lower than the standard's, and the diluted concentration is lower than the standard's, as it should be.

How it connects

  • Back to 3.7: molarity, and the dilution relation M1V1 = M2V2, are used to undo the dilution here.
  • Back to 3.12: the choice of wavelength rests on the idea that a substance absorbs photons whose energies match its energy gaps.
  • Forward to 5.1: spectrophotometry is a common way to follow the changing concentration of a colored species during a reaction.
  • Forward to 7.4: equilibrium concentrations of colored species are often found from absorbance.

Common mistakes

  • Forgetting the dilution. If the measured sample was diluted, multiply by the dilution factor to get the original concentration.
  • Using a wavelength other than the maximum. The line becomes shallower and the measurement less sensitive.
  • Reading outside the calibration range. Extending the line past the standards is not supported by data.
  • Treating absorbance as having units. A is unitless, and the units belong to epsilon.
  • Ignoring smudges and leftover liquid. Fingerprints raise absorbance, and water left in an unrinsed cuvette dilutes the sample, so absorbance and the calculated concentration come out too low.

How it shows up on the exam

  • A table of absorbance and concentration values for standards is given, and you find an unknown or the molar absorptivity from the slope.
  • A lab procedure is described and you identify a step that would cause an error, and say how it changes the result.
  • You explain why a particular wavelength is chosen or why a graph of absorbance against concentration should be linear.

Check yourself

  1. If a solution of concentration c gives A = 0.30, what absorbance do you predict at 2c in the same cuvette?
  2. Why do you choose the wavelength of maximum absorbance before preparing standards?
  3. A fingerprint is left on the cuvette. How does the reading change, and how would the calculated concentration be affected?
  4. Why should an unknown fall within the range of the standards?

Practice this topic. AP Practice questions on exactly what this lesson covers, with an explanation for every answer.